Automatic calibration system (data sheet 10465)

Absolute, gauge or differential pressure

10465 automatic calibration system DPC4800

model DPC 4800
version up to 3 pressure ranges and barometric reference
accuracy DPC 4800 A (all-rounder)
measurement uncertainty (k = 2): 0.02 % FS + 0.02 % rdg
control accuracy: 0.015 %

DPC 4800 A+ (all-rounder plus)
measurement uncertainty (k = 2): 0.008 % FS + 0.006 % rdg
control accuracy: 0.008 %

DPC 4800 P (precision)
measurement uncertainty (k = 2): 0.006 % FS + 0.003 % rdg
control accuracy: 0.005 % (optional 0.003 %)
medium gas
pressure ranges overpressure: −1 / +1 bar to 0 – 230 bar (−30 inHg / +15 psi to 0 – 3000 psi)
absolute pressure: 0 – 1 bar to 0 – 100 bar (0 – 15 psi to 0 – 1500 psi)
differential pressure: ±30 mbar to ±300 mbar (±1 inHg to ±10 inHg)
specialty
  • controlling and measuring test pressures with a single device
  • up to 3 precision sensors can be actuated automatically (plus barometric reference)
  • customised configurations of the pressure controller possible
  • very high measuring rate
  • very large control volumes possible
  • large colour touch screen, LED backlight
  • modern interfaces for PLC automation environments
  • easily calibrated
  • modular construction
  • fully digital measuring instrument
  • automatic creation of test certificates via full version calibration software DynaCal
application
  • transfer standard for calibration laboratories
  • high-precision pressure source and pressure reference for the construction of test benches
  • actuator and reference in modern PLC topologies
  • laboratories for research and development
  • production means in pressure gauge, pressure switch and sensor manufacturing
  • tool for transmitter calibration and creation of certificates
  • long-term measurements