Resistance thermometers (data sheet 8526)

Intrinsically safe


The resistance thermometer models TPtHrXiA and TPtHrXiAT are designed for the installation into solid drilled or fabricated thermometer thermowells (e.g. according to DIN 43 772). Without thermowell, this version must not be applied in pressurised media.
They have an EU Type Examination Certificate for the ignition protection type “intrinsic safety” and meet the requirements of the directive 2014/34/EU for the application in potentially explosive atmospheres due to gases and dusts.
For both models, we offer various measuring resistors according to DIN EN 60 751. In addition, model TPtHrXiAT is available with several fitted transmitters with analogue or digital output.
 
8526 Resistance thermometers TPtHrXiA intrinsically safe for the installation into thermowells measuring range -200 °C to +600 °C (-328 °F to 1112 °F) ARMANO

model TPtHrXiA / TPtHrXiAT
operating temperature range −200 / +600 °C (−328 / +1112 °F)
measuring insert replaceable
connection heads types XE-BUZ, XE-BUZ-H, XI-BUZ or XI-BUZ-H
degree of protection (DIN EN 60 529) IP67 when mounted in a thermowell and corresponding screwed cable gland (PAN, PAR, MAN, MAR)
approvals II 1G Ex ia IIC T6…T1 Ga
II 1/2G Ex ia IIC T6…T1 Ga/Gb1)
II 2G Ex ia IIC T6…T1 Gb
II 1D Ex ia IIIC T80 °C…T440 °C Da
II 1/2D Ex ia IIIC T80 °C…T440 °C Da/Db1)
II 2D Ex ia IIIC T80 °C…T440 °C Db

1) only with thermowell suitable for zone separation – see operating instructions B71
application in explosion hazardous areas zones 0, 1, 2, 20, 21, 22
output signal model TPtHrXiA:   resistance according to DIN EN 60 751
model TPtHrXiAT: 4…20 mA, HART® or PROFIBUS® PA / FOUNDATION™ Fieldbus